Assessment of simpler calibration models in the development and validation of a fast postmortem multi-analyte LC-QTOF quantitation method in whole blood with simultaneous screening capabilities using SWATH acquisition

Marco P. Elmiger, Thomas Kraemer, Andrea E. Steuer, Michael Poetzsch

Abstract

In postmortem toxicology, fast methods can provide a triage to avoid unnecessary autopsies. Usually, this requires multiple qualitative and quantitative analytical methods. The aim of the present study was to develop a postmortem LC-QTOF method for simultaneous screening and quantitation using easy sample preparation and reduced alternative calibration models. Hence, a method for 24 highly relevant substances in forensic toxicology was fully validated using the following calibration models: one-point external, one-point internal via corresponding deuterated standards, multi-point external daily calibration, and multi-point external weekly calibration. Two hundred microliters of postmortem blood were spiked with internal deuterated standard mixture and extracted by acetonitrile protein precipitation. Analysis was performed on a Sciex 6600 QTOF instrument with ESI+ mode using data-independent acquisition (DIA) namely sequential window acquisition of all theoretical mass spectra (SWATH). Validation of the different calibration models included selectivity, autosampler stability, recovery, matrix effects, accuracy, and precision for 24 substances. In addition, corresponding deuterated analogs of 52 substances were included to the internal standard mix for semi-quantitative concentration assessment. The simple protein precipitation provided recoveries higher than 55 and 75% for all analytes at low and high concentrations, respectively. Accuracy and precision criteria (bias and imprecision ± 15 and ± 20% near the limit of quantitation) were fulfilled by the different calibration models for most analytes. The validated method was successfully applied to more than 100 authentic postmortem samples and 3 proficiency tests. Furthermore, the one-point internal calibration via corresponding deuterated standard proved to be a considerably time saving technique for 76 analytes.

Graphical abstract

One-point and multi-point calibration and the resulting beta-tolerance intervals from method validation

Publisher URL: https://link.springer.com/article/10.1007/s00216-017-0594-1

DOI: 10.1007/s00216-017-0594-1

You might also like
Discover & Discuss Important Research

Keeping up-to-date with research can feel impossible, with papers being published faster than you'll ever be able to read them. That's where Researcher comes in: we're simplifying discovery and making important discussions happen. With over 19,000 sources, including peer-reviewed journals, preprints, blogs, universities, podcasts and Live events across 10 research areas, you'll never miss what's important to you. It's like social media, but better. Oh, and we should mention - it's free.

  • Download from Google Play
  • Download from App Store
  • Download from AppInChina

Researcher displays publicly available abstracts and doesn’t host any full article content. If the content is open access, we will direct clicks from the abstracts to the publisher website and display the PDF copy on our platform. Clicks to view the full text will be directed to the publisher website, where only users with subscriptions or access through their institution are able to view the full article.