VARIABILITY IN FOVEAL AVASCULAR ZONE AND CAPILLARY DENSITY USING OPTICAL COHERENCE TOMOGRAPHY ANGIOGRAPHY MACHINES IN HEALTHY EYES

VARIABILITY IN FOVEAL AVASCULAR ZONE AND CAPILLARY DENSITY USING OPTICAL COHERENCE TOMOGRAPHY ANGIOGRAPHY MACHINES IN HEALTHY EYES
Ferenczy, Sandor, Sioufi, Kareem, Magrath, George N., Shields, Carol L., Samara, Wasim A., Say, Emil Anthony T.
Purpose:To evaluate the variability in foveal avascular zone (FAZ) and capillary density measurements on optical coherence tomography angiography using Optovue RTVue XR Avanti (OA) (Optovue) and Zeiss Cirrus HD-OCT 5000 (ZC) (Carl Zeiss Meditec). Methods:In this prospective, comparative case series, parafoveal (3 × 3 mm) optical coherence tomography angiography scans were obtained on healthy volunteers using both the Avanti and Cirrus. The FAZ area and capillary density at the level of both the superficial and deep capillary plexus were measured automatically using the built-in ReVue software (Optovue) with the Avanti as well as manually using ImageJ (National Institutes of Health) with both machines. Results:There were 50 eyes in 25 healthy volunteers included in the analysis. Mean subject age was 33 years and there were 14 women (56%). On optical coherence tomography, mean central macular thickness was significantly greater on OA (259.1 μm) than ZC (257.6 μm, P = 0.0228). On optical coherence tomography angiography, mean superficial and deep plexus FAZ measured 0.2855 mm2 and 0.3465 mm2 on Avanti automated (A-A), 0.2739 mm2 and 0.3637 mm2 on Avanti manual (A-M), and 0.2657 mm2 and 0.3993 mm2 on Cirrus manual (C-M), respectively. There were no statistically significant differences in superficial plexus FAZ measurements between the A-A and A-M (P = 0.4019) or A-A and C-M (P = 0.1336). The A-M measured significantly larger than C-M (P = 0.0396). Deep plexus FAZ measurements were similar on A-A and A-M (P = 0.6299), but both were significantly less compared with C-M (P < 0.0001 for A-A vs. C-M, P = 0.0184 for A-M vs. C-M). Mean superficial and deep plexus capillary densities were 53.6% and 59.3% on A-A, 48.1% and 47.7% on A-M, and 52.5% and 48.1% on C-M, respectively. Superficial plexus capillary density measurements were statistically similar on A-A and C-M (P = 0.0623), but both were significantly higher than A-M (P < 0.0001 for A-A vs. A-M, P < 0.0001 for A-M vs. C-M). However, deep plexus capillary density measurements on A-A were significantly higher than A-M (P < 0.0001) and C-M (P < 0.0001), but A-M and C-M measurements were similar (P = 0.5986). There was no significant difference in all parameters measured in both eyes of one subject using any of the three measuring techniques. Conclusion:While measurements taken with the same machine and technique are consistent and reliable between fellow eyes, significant variability exists in FAZ and capillary density measurements among different machines and techniques. Comparison of measurements across machines and techniques should be considered with caution.
You might also like
Discover & Discuss Important Research

Keeping up-to-date with research can feel impossible, with papers being published faster than you'll ever be able to read them. That's where Researcher comes in: we're simplifying discovery and making important discussions happen. With over 19,000 sources, including peer-reviewed journals, preprints, blogs, universities, podcasts and Live events across 10 research areas, you'll never miss what's important to you. It's like social media, but better. Oh, and we should mention - it's free.

  • Download from Google Play
  • Download from App Store
  • Download from AppInChina

Researcher displays publicly available abstracts and doesn’t host any full article content. If the content is open access, we will direct clicks from the abstracts to the publisher website and display the PDF copy on our platform. Clicks to view the full text will be directed to the publisher website, where only users with subscriptions or access through their institution are able to view the full article.