5 years ago

Adjustable loop ACL suspension devices demonstrate less reliability in terms of reproducibility and irreversible displacement

Atesch Ateschrang, Takura Mukabeta, Benjamin Voumard, Sandro Kohl, Dimitrios S. Evangelopoulos, Sufian S. Ahmad, Philippe Zysset, Michael T. Hirschmann

Abstract

Purpose

The aim of this study was to perform a comprehensive biomechanical examination of frequently applied femoral cortical suspension devices, comparing the properties of both fixed and adjustable fixation mechanisms. It was hypothesized that adjustable loop devices demonstrate less consistent fixation properties with increased variability compared to fixed loop devices.

Methods

Nine frequently applied fixation button types were tested, six adjustable and three rigid loop devices. Six samples of each device type were purchased. Each device was installed in a servo-hydraulic mechanical testing machine, running a 2000 cycle loading protocol at force increments between 50 and 500 N. Irreversible displacement in mm was measured for all of the tested samples of each implant. Ultimately, maximum load to failure was applied and measured in Nm. An irreversible displacement of 3 mm was considered failure of the implant.

Results

Three of the six adjustable devices (GraftMax™, TightRope® ToggleLoc™) demonstrated a median displacement above the threshold of clinical failure before completion of the cycles. All adjustable loop devices showed a wide intragroup variation in terms of irreversible displacement, compared to fixed-loop devices. Fixed-loop devices provided consistent reproducible results with narrow ranges and significantly lower irreversible displacement (p < 0.05), the maximum being 1.4 mm. All devices withstood an ultimate force of more than 500 N.

Conclusion

Adjustable loop devices still show biomechanical inferiority and demonstrate heterogeneity of fixation properties with wide- and less-reproducible displacement ranges resultant to the mechanism of adjustment, denoting less reliability. However, three adjustable devices (RIGIDLOOP™ Adjustable, Ultrabutton , ProCinch™) demonstrate fixation capacities within the margins of clinical acceptance. RIGIDLOOP™ Adjustable provides the most comparable fixation properties to fixed loop devices.

Publisher URL: https://link.springer.com/article/10.1007/s00167-017-4769-2

DOI: 10.1007/s00167-017-4769-2

You might also like
Discover & Discuss Important Research

Keeping up-to-date with research can feel impossible, with papers being published faster than you'll ever be able to read them. That's where Researcher comes in: we're simplifying discovery and making important discussions happen. With over 19,000 sources, including peer-reviewed journals, preprints, blogs, universities, podcasts and Live events across 10 research areas, you'll never miss what's important to you. It's like social media, but better. Oh, and we should mention - it's free.

  • Download from Google Play
  • Download from App Store
  • Download from AppInChina

Researcher displays publicly available abstracts and doesn’t host any full article content. If the content is open access, we will direct clicks from the abstracts to the publisher website and display the PDF copy on our platform. Clicks to view the full text will be directed to the publisher website, where only users with subscriptions or access through their institution are able to view the full article.